Command
Description
Command Syntax
(command underlined)
ACCUBUBBLE response
(underlined)
"a" represents the single-
character address
XPA Bubbler
reading
averaging
aXPA+nspeed!
Nsamples is the number of samples
to average. If speed is 0 or omitted
then the AccuBubble operates in the
default more accurate slow mode. If
speed is non-zero the AccuBubble
will operate in fast mode. Preferred
over the aXT command for setting up
averaging when operating in a
bubbler mode.
a0012
Indicating the command will be
complete in 1 second, returning two
values.
a
Service request
acs
Where c is the number of samples to be
averaged and s indicates the speed where
0 is slow and 1 is fast.
XPC
Suggests
bubbler pump
and read
times
aXPC+minpurge!
If a value for minpurge is given, then
the pump will be run for at least
minpurge seconds before attempting
to determine purge_off otherwise it
will run for at least 10 seconds.
aD0!
a2553
Indicating the command will be
complete in 255 seconds, returning three
values.
a
service request
apos
Where p is the purge_on time detected, o
is the purge_off time detected, and s is
the pump_off time detected for currently
programmed pump_on time.