Command
Description
Command Syntax
(command underlined)
ACCUBAR
®
response (underlined)
"a" represents the
single-character address
XUP Set pressure
units
(Version 1.5
and higher for
setting
number of
decimal
places)
aXUP+n+d!
n = 0 mB (hPa)
n = 1 "Hg
n = 2 kPa
n = 3 mmHg
n = 4 Atm
n = 5 psia
n = 9 user units
d = number of places right of the
decimal (optional parameter)
Example: 0XUP+9+2!
Select user units with 2 right digits--
(make sure you use XUU command
to set the scale and offset for the
desired user units)
a0012
indicating the command will be complete
in 1 second and 2 values can be collected
Note:
a D0 command issued after the
XUP will return the value of the units
that are selected and the number of digits
right of the decimal point.
Note Version 2.0 and higher:
Current
settings can be queried by issuing the
aXUP! Command followed by the aD0!
Command.
XUT Set
temperature
units
aXUTn!
n = 0 for Celsius,
n = 1 for Fahrenheit
Example: 0XUT1!
(set temperature units to F)
Note Version 2.0 and higher:
Current settings can be queried by
issuing the aXUT! Command
followed by the aD0! Command.
a0011
indicating the command will take 1
second to complete and 1 value can be
collected.
Note:
a D0 command issued after the
XUT will return the value of the units
that are selected.
XUU Set User
Units
aXUUso!
where s is the pressure scale factor
and o is the offset,
User output = (mB)*scale + offset
Example: 0XUU+27.63+0
(27.63 user units per mB)
Note Version 2.0 and higher:
Current settings can be queried by
issuing the aXUU! Command
followed by the aD0! Command.
a0012
Note:
a D0 command issued after XUU
will return the scale and offset.
Note
: a scale of 0 is invalid.
Note:
Be sure that the units of pressure
(XUP) are set to user units (9).