Bringing the Benefits of Real-Time Data Collection to the World
Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164-4444
29
Command
Description
Command Syntax
(command underlined)
ACCUBAR
®
response (underlined)
"a" represents the
single-character address
M1
Request
Pressure
Measurement
in mB
(factory
calibrated
value)
aM1!
aD0!
attt1
ttt is the time in seconds until the
measurement is ready, 1 is the number of
values that can be collected.
a
service request
ap
where p is the signed pressure value in
mB
M2
Request
Temperature
Measurement
aM2!
aD0!
attt2
ttt is the time in seconds until the
measurement is ready and 2 is the
number of values that can be collected
a
service request
atu
where t is the temperature and u is the
units 0= Celsius and 1=Fahrenheit. Use
the XUT command to set the units.
M3
Request User
Scale, User
Offset, and
Field
Calibration
Offset
aM3!
aD0!
a0003
000 is the time in seconds until the
measurement is ready and 3 is the
number of values that can be collected
asoc
where, s is the user scale and o is the
user offset, and c is the field calibration
offset. The field calibration is returned in
the current units.
M4
Request
Standards lab
Calibration
Scale and
Offset
aM4!
aD0!
a0002
000 is the time in seconds until the
measurement is ready and 2 is the
number of values that can be collected
aso
where, s is the scale calibration and o is
the offset calibration(mB).