32
Command
Description
Command Syntax
(command underlined)
Sensor response (underlined)
"a" represents the
single-character address
MC6 Request
Temperature
and Pressure
Measurement
with CRC-16
(Version 2.0
and higher)
aMC6!
aD0!
attt04
ttt is the time in seconds until the
measurement is ready and 04 is the
number of values that can be collected
a
service request
atupvC
where t is the temperature, u is the
temperature units, p is the pressure, v is
the pressure units, and C is the 3
character CRC. Use the XUT command
to set the temperature units and the XUP
command to set the pressure units.
MC7 Request
factory
calibration
Pressure and
Temperature
Measurement
with CRC-16
(Version 2.0
and higher)
aMC7!
aD0!
attt02
ttt is the time in seconds until the
measurement is ready and 02 is the
number of values that can be collected
a
service request
aptC
where p is the pressure in mB, t is the
temperature in degrees Celsius, and C is
3 character CRC.
R0
R1
.
.
.
R9
Request
Continuous
Measurement
Readings
aR0!
aR1!
.
.
.
aR9!
a
Unit only returns its address because it
does not support continuous
measurements
.
V
Initiate Verify
sequence
aV!
aD0!
attt5
indicating that the command will be
complete in ttt seconds and 5 values can
be collected.
a
service request
arespu
where r is the ROM checksum, e is the
EEROM checksum, s is the number of
resets since power up, p is the number of
power ups, and u is the number of
unexpected interrupts.