Bringing the Benefits of Real-Time Data Collection to the World
Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164-4444
31
Command
Description
Command Syntax
(command underlined)
Sensor response (underlined)
"a" represents the
single-character address
MC2 Request
Temperature
Measurement
with CRC-16
(Version 2.0
and higher)
aMC2!
aD0!
attt2
ttt is the time in seconds until the
measurement is ready and 2 is the
number of values that can be collected
a
service request
atuC
where t is the temperature, u is the units
0= Celsius and 1=Fahrenheit, and C is
the 3 character CRC. Use the XUT
command to set the units.
MC3 Request User
Scale, User
Offset, and
Field
Calibration
Offset with
CRC-16
(Version 2.0
and higher)
aMC3!
aD0!
a0003
000 is the time in seconds until the
measurement is ready and 3 is the
number of values that can be collected
asocC
where, s is the user scale and o is the
user offset (mB), c is the field calibration
offset (current pressure units), and C is
the 3 character CRC.
MC4 Request
Standards lab
Calibration
Scale and
Offset with
CRC-16
(Version 2.0
and higher)
aMC4!
aD0!
a0002
000 is the time in seconds until the
measurement is ready and 2 is the
number of values that can be collected
asoC
where, s is the scale calibration, o is the
offset calibration (mB), and C is the 3
character CRC.