8.5.2 NaCl Standard Solution
total 2 standard solutions including 10.683mS(0.1 mol/L) and 251.3mS(saturated) can be
selected.
Please refer to “Appendix Calibration Solution: Conductivity and Temperature Table” for your
reference.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
ATC Mode
MTC Mode
Press or to select
standard solutions from 10.683mS or
251.3mS
Press or to
calibrate the NaCl solution
Press to confirm it and start calibration.
Press or to adjust
the standard solution value.
Press to confirm it.
Under MTC temperature
mode, press or
to adjust temperature
value.
8