Function Code: 01H & 05H Modbus Response (Dispersion Parameter)
Logic
address
R/W Item BIT
Description Default
value Note
0070H R
LO Alarm
1
Contact on
0 (Contact off)
0071H R
Hi Alarm
1
Contact on
0 (Contact off)
0072H R
MA Too High
1
Contact on
0 (Contact off)
0073H R
MATtoo Low
1
Contact on
0 (Contact off)
0074H R Exceed Temp. Range
1
Contact on
0 (Contact off)
0075H R
Exceed
Measuring Range
1
Contact on
0 (Contact off)
0076H R
RLY1 Action
1
Contact on
0 (Contact off)
0077H R
RLY2 Action
1
Contact on
0 (Contact off)
0078H R
Clean Action
1
Contact on
0 (Contact off)
0079H-
008FH
Factory reserved
рй
Summary of Contents for EC-4110-I
Page 1: ...1 EC 4110 I EC 4110 ICON Intelligent Conductivity Transmitter Operation Manual...
Page 78: ...11 Installation of Cells 11 1 Appearance of cells 8 201 PFA unit mm...
Page 84: ...31 11 995 282 7 32 12 220 288 33 12 445 293 3 34 12 673 298 7 35 12 902 304 1 36 13 132 309 5...
Page 85: ......
Page 86: ......