![Suntex EC-4110-I Operation Manual Download Page 82](http://html1.mh-extra.com/html/suntex/ec-4110-i/ec-4110-i_operation-manual_1400958082.webp)
Appendix: Calibration Solution
Potassium chloride solution
Conductivity in mS/cm
䉝
Conductivity
1.413mS
烏
25
䉝
0.01mol/l
12.88mS
烏
25
䉝
0.1mol/l
111.8mS
烏
25
䉝
1mol/l
0
0.776 7.15
65.41
5
0.896 8.22
74.14
10
1.020 9.33
83.19
15
1.147 10.48
95.52
16
1.173 10.72
94.41
17
1.199 10.95
96.31
18
1.225 11.19
98.22
19
1.251 11.43 100.14
20
1.278 11.67 102.07
21
1.305 11.91
104
22
1.332 12.15 105.94
23
1.359 12.39 107.89
24
1.386 12.64 109.84
25
1.413 12.88
111.8
26
1.441 13.13 113.77
27
1.468 13.37 115.74
28
1.496 13.62
29
1.524 13.87
30
1.552 14.12
31
1.581 14.37
32
1.609 14.62
33
1.638 14.88
34
1.667 15.13
35
1.696 15.39
36
15.64
рр