Temperature Coefficient
Linear temperature compensation from (0.00%~ 40.00%),
Non-Linear compensation, or No compensation
Compensation Temperature
Freely set reference temperature 0~200
Û
C
Display Screen
Large LCM with sensitization sensor for auto/manual illumination
function and contract function
Languages
English / Traditional Chinese / Simplified Chinese
Display
Text mode: Numerical display
Chart mode: 3 mins real-time dynamic graph
(conc. and cond. only)
Trace mode: Set up from 3 mins to four weeks duration of the
measured value trend graph
(conc. and cond. only)
Analog output
Isolated DC 0/4~20mA corresponding to main measurement,
max. load 500
:
Logbook
50 event records
RS-485 interface
Isolated (MODBUS RTU or ASCII)
Modbus communication
Offer measuring value, calibration data, product adjustment,
logbook, control parameter, etc or system setup
Settings
Contact
RELAY contact, 240VAC 0.5A Max.(recommended)
Activate
Hi/Lo. Hi/Hi. Lo/Lo selectable two limited programmable,
ON/OFF
Wash
RELAY contact ON 0~99min. 59sec. / OFF 0~999hr 59min.
Protection
IP65 (NEMA 4X)
Power Supply
100V~240VAC±10%, 7W max., 50/60Hz
Installation
Wall or Pipe or Panel Mounting
Dimensions
96m × 96mm × 132mm (H
u
W
u
D)
Cut off Dimensions
93 mm
u
93 mm (H
u
W)
Weight
0.5Kg
Note: The specifications are subject to change without notice.
о
Summary of Contents for EC-4110-I
Page 1: ...1 EC 4110 I EC 4110 ICON Intelligent Conductivity Transmitter Operation Manual...
Page 78: ...11 Installation of Cells 11 1 Appearance of cells 8 201 PFA unit mm...
Page 84: ...31 11 995 282 7 32 12 220 288 33 12 445 293 3 34 12 673 298 7 35 12 902 304 1 36 13 132 309 5...
Page 85: ......
Page 86: ......
Page 87: ......