7.11 Current Output
Enter setup of Current Output. Select 0~20mA or 4~20mA current output. Set the related value
to the range of Cond. measurement. If the range of the Cond. measurement is to be set smaller,
the resolution of current output is higher. When the measured value exceeds the higher range
limit, the current will remain approximately 22mA output. When the measured value exceeds
the lower range limit, under 0~20mA mode the current output will remain 0mA output; while
under 4~20mA mode the current output will remain approximately 2mA output.
The
exceptional output value can be used as a basis for failure determination.
Under
HOLD(measurement) status, the current output maintain the last output value before HOLD
status. However, in order for convenience of insuring the current setting of an external recorder
or of a PLC controller, the current output will be 0/4mA or 20mA under the analog output setup
menu.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press
ġ
or
to select to current
output range,
0-20mA or 4-20mA.
Press or to set
the lower limit of Cond.
value relative to 4.0mA
o u t p u t s i g n a l . I f t h e
m e a s u r i n g r a n g e o f
conductivity measurement
is set as AUTO, then it
needs to
EHVHWXQLWȝ6FP
or mS/cm), value, and
decimal point by orders.
Press or to set
the lower limit of Cond.
value relative to 0.0mA
o u t p u t s i g n a l . I f t h e
m e a s u r i n g r a n g e o f
conductivity measurement
is set as AUTO, then it
QHHGVWREHVHWXQLWȝ6
/cm
or mS/cm), value, and
decimal point by orders.
нп
Summary of Contents for EC-4110-I
Page 1: ...1 EC 4110 I EC 4110 ICON Intelligent Conductivity Transmitter Operation Manual...
Page 78: ...11 Installation of Cells 11 1 Appearance of cells 8 201 PFA unit mm...
Page 84: ...31 11 995 282 7 32 12 220 288 33 12 445 293 3 34 12 673 298 7 35 12 902 304 1 36 13 132 309 5...
Page 85: ......
Page 86: ......
Page 87: ......