7.5 Product Adjustment
Enter setup of Product Adjustment to make the fine adjustment of the measurement reading.
There is no need for users to take out the sensor from the field and calibrate it. Via this function
the measuring values from the field can be compared to the result that came from the sampling
and directly adjusted to the same value. If turning on this function, there shows the symbol
“PDT” on the display screen.(please refer to ch6.4 “Warning symbols and text”).
The inductive sensor applies electromagnetic effects to measure the conductivity of a sample
solution which can be affected by length of signal cable and the distance between a sensor and
the shell of pipe, making an influence on the cell factor and measuring value. At the moment,
Product Adjustment can be used to calibrate the measuring value without changing the cell
factor.
Users may compare the reading with the
actual measured value. Press or
to adjust the Cell Factor( a cell constant’s
multiplier) in the main display in order to
modify the reading of measurement
which shows in the bottom part of
display.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Enter
Ⱦ
Temperature
ȿ
Setup
Press or
to switch on or
switch off.
мт
Summary of Contents for EC-4110-I
Page 1: ...1 EC 4110 I EC 4110 ICON Intelligent Conductivity Transmitter Operation Manual...
Page 78: ...11 Installation of Cells 11 1 Appearance of cells 8 201 PFA unit mm...
Page 84: ...31 11 995 282 7 32 12 220 288 33 12 445 293 3 34 12 673 298 7 35 12 902 304 1 36 13 132 309 5...
Page 85: ......
Page 86: ......
Page 87: ......