Torq-Tronics 2
®
/
Owners Manual / Appendix A: Serial Commands
A.4
Request tester to generate and print a report of data captured
A.4.1 Type: DR<Enter>
Data report structure depends on settings of the print flags, see TX command.
Example of a DR<Enter> command, dump report.
Start Report Tester Serial number: 00001234,
Controlled by PSerialNum
Tester1234567890,
Controlled by PUserString
Records
in
Report:
004,
Controlled by PRecInRpt
+41.2597,In.oz,PEAK,NA,001,00001234,Test Station 1,.000001,.000001,05/31/2011,11:45:34
See above TX command
+14.8275,In.oz,PEAK,NA,002,00001234,Test Station 1,.000001,.000001,05/31/2011,11:55:55
TX command
+14.3511,In.oz,PEAK,NA,003,00001234,Test Station 1,.000001,.000001,05/31/2011,11:56:10
TX command
+14.2901,In.oz,PEAK,NA,004,00001234,,Test Station 1,.000001,.000001,05/31/2011,11:56:27
TX command
End Report Tester Serial number: 00001234,
Controlled by PSerialNum
A.5
Serial Messaging Command List
A.5.1 Special ASCII commands:
A.5.1.1 BEL
= 0x0007 ASCII Bell. This is used to indicate invalid data in message.
A.5.1.2 CR
= 0x000D ASCII Carriage return
A.5.1.3 LF
= 0x000A ASCII Line Feed
A.5.1.4 <CR+LF> = <Enter > = ASCII Carriage Line Feed.
A.5.1.5 Enter Calibration Mode type: CM< Enter>
Returns message: CalOn<CR+LF>
A.5.1.6 Exit calibration mode and unlock the keypad type: CMD< Enter >
Returns message: CalOff<CR+LF>
A.5.1.7 Toggles the Enter/exit state of Diagnostic mode type: D< Enter >
Depending on the state, returns: DiaOn<CR+LF> or DiaOff<CR+LF>
A.5.1.8 Enter Diagnostic mode Type: D1< Enter >
Returns message: DiaOn<CR+LF>
A.5.1.9 Exit Diagnostic mode: Type D0< Enter > Note: the 0 is a ZERO.
Returns message: DiaOff<CR+LF>
Owners Manual
Torq-Tronics 2
Sturtevant Richmont
Direct: 1 847/455-8677
Toll-Free: 800/877-1347
www.srtorque.com
Example of a DR<Enter> command, dump report.
Start Report Tester Serial number: 00001234,
Controlled by PSerialNum
Tester1234567890,
Controlled by PUserString
Records in Report: 004,
Controlled by PRecInRpt
+41.2597,In.oz,PEAK,NA,001,00001234,Test Station 1,.000001,.000001,05/31/2011,11:45:34
See above TX command
+14.8275,In.oz,PEAK,NA,002,00001234,Test Station 1,.000001,.000001,05/31/2011,11:55:55
TX command
+14.3511,In.oz,PEAK,NA,003,00001234,Test Station 1,.000001,.000001,05/31/2011,11:56:10
TX command
+14.2901,In.oz,PEAK,NA,004,00001234,,Test Station 1,.000001,.000001,05/31/2011,11:56:27
TX command
End Report Tester Serial number: 00001234,
Controlled by PSerialNum
11
/16/201
7
REV
E
. Document Page 29 of 39