Torq-Tronics 2
® /
Owners Manual
16.4.49 Again hang the weight platform (hook) and the appropriate weights (20%) (found in
the “Calibration Chart”) on the calibration arm-gently, and imparting as little motion
to the weights as possible. Wait for the arm to completely stop moving then press the
‘CLEAR’ or ‘DOWN’
key on the unit.
Observe acknowledgement of the 80% counter clockwise calibration point.
16.4.50 Press the ‘MENU’ or ‘UP’ key to move to the next test point.
16.4.51 Again hang the weight platform (hook) and the appropriate weights (30%) (found in
the “Calibration Chart”) on the calibration arm-gently, and imparting as little motion
to the weights as possible. Wait for the arm to completely stop moving then press the
‘CLEAR’ or ‘DOWN’
key on the unit.
Observe acknowledgement of the 30% counter clockwise calibration point.
16.4.52 Press the ‘MENU’ or ‘UP’ key to move to the next test point.
16.4.53 Again hang the weight platform (hook) and the appropriate weights (40%) (found in
the “Calibration Chart”) on the calibration arm-gently, and imparting as little motion
to the weights as possible. Wait for the arm to completely stop moving then press the
‘CLEAR’ or ‘DOWN’
key on the unit.
Observe acknowledgement of the 40% counter clockwise calibration point.
16.4.54 Press the ‘MENU’ or ‘UP’ key to move to the next test point.
16.4.55 Again hang the weight platform (hook) and the appropriate weights (50%) (found in
the “Calibration Chart”) on the calibration arm-gently, and imparting as little motion
to the weights as possible. Wait for the arm to completely stop moving then press the
‘CLEAR’ or ‘DOWN’
key on the unit.
Observe acknowledgement of the 50% counter clockwise calibration point.
16.4.56 Press the ‘MENU’ or ‘UP’ key to move to the next test point.
16.4.57 Again hang the weight platform (hook) and the appropriate weights (60%) (found in
the “Calibration Chart”) on the calibration arm-gently, and imparting as little motion
to the weights as possible. Wait for the arm to completely stop moving then press the
‘CLEAR’ or ‘DOWN’
key on the unit.
Observe acknowledgement of the 60% counter clockwise calibration point.
16.4.58 Press the ‘MENU’ or ‘UP’ key to move to the next test point.
16.4.59 Again hang the weight platform (hook) and the appropriate weights (70%) (found in
the “Calibration Chart”) on the calibration arm-gently, and imparting as little motion
to the weights as possible. Wait for the arm to completely stop moving then press the
‘CLEAR’ or ‘DOWN’
key on the unit.
Observe acknowledgement of the 70% counter clockwise calibration point.
16.4.60 Press the ‘MENU’ or ‘UP’ key to move to the next test point.
16.4.61 Again hang the weight platform (hook) and the appropriate weights (80%) (found in the
“Calibration Chart”) on the calibration arm-gently, and imparting as little motion to the
weights as possible. Wait for the arm to completely stop moving then press the ‘CLEAR’ or
‘DOWN’
key on the unit.
Observe acknowledgement of the 80% counter clockwise calibration point.
16.4.62 Press the ‘MENU’ or ‘UP’ key to move to the next test point.
16.4.63 Again hang the weight platform (hook) and the appropriate weights (90%) (found in the
“Calibration Chart”) on the calibration arm-gently, and imparting as little motion to the
weights as possible. Wait for the arm to completely stop moving then press the ‘CLEAR’ or
‘DOWN’
key on the unit.
Observe acknowledgement of the 90% counter clockwise calibration point.
Owners Manual
Torq-Tronics 2
Sturtevant Richmont
Direct: 1 847/455-8677
U.S. Toll-Free: 800/877-1347
www.srtorque.com
11/16/201
7
REV
E
. Document Page 25 of 39
Note:
Torque Tool Manager 2 is a
for purchase software that creates a
torque testing and calibration histo-
ry database. Torque Tool Manager
works with the legacy Torq-Tronics
unit as well as System 4/5. Torque
Tool Manager 2 software function-
ality is not as robust with the new
Torq-Tronics 2 unit.
Torque Tool Manager software is
sold separately. For more infor-
mation email customerservice@
srtorque.com.