Torq-Tronics 2
® /
Owners Manual
8.3
General Set Up Process For All Tests:
The following process of setting up Torq-Tronics 2 for testing can be used to verify or
calibrate any acceptable tool type. (Impact tools are not acceptable.)
8.4 Memory:
If you see three digits in the lower right hand side of the display, memory function is set to
on. If you do not see three digits in the lower right hand side of the Display the memory
function is set to off.
The factory default setting for the memory function is “OFF.” To turn the memory on,
press MENU and then press the UP arrow and then press the MENU button and memory
will be activated.
If the memory is on and you would like to turn it off, press the MENU and then ZERO
button. Then press the MENU button to return to normal functioning.
8.5
Clearing The Memory:
The Torq-Tronics 2 memory can be cleared the following ways.
8.5.1
The operator may choose to download a report by pressing the MENU button – the display
will read Data and the number of tests or records currently resident in the unit. The opera-
tor then uses the DOWN ARROW to download the report to Hyper-Terminal via RS232
port. When the download is complete the lower left hand side of the display shows CLR
which provides the order to press the CLEAR button. Press the CLEAR button to clear
the memory. You will see the counter in the lower right hand of the display move from the
number of records that were in the unit down to 001. The display always displays the cur-
rent record unless you are in “VR” (view record) mode.
8.5.2
The unit memory may be cleared without printing by MENU button and then pressing the
CLEAR button.
8.5.3
Accessing Torq-Tronics 2 via Hyper-Terminal allows the operator to clear the memory by
using the serial command CL<CR+LF> This returns the message DONE<CL+LF> upon
task completion.
8.5.4
The unit holds 999 records. If the records are not cleared, after the 999th test, the unit will
start to overwrite the subsequent records. The records will retain their numbering sequence,
but the data from the original test 001 will be replaced by the data from test 1000.
8.6
Here is the sequence to use when setting up tests in Torq-Tronics 2.
8.6.1
Select the mode by pressing the PEAK TORQUE key and then using the up arrow to select
the appropriate test mode. When you have selected the desired test mode press the PEAK
TORQUE key again. If you have selected Power Tool mode, use the down arrow to select
the appropriate filter. Once you have the set the filter, press the PEAK TORQUE key again.
8.6.2. To set the units of measure press the UNITS and use the UP ARROW to scroll through
the listing until you find the unit of measure you want. The unit of measure is located on
the display in the upper right hand corner. When you have scrolled through and found
the desired unit of measure, stop. At that point press the UNITS button again to return to
normal functioning. Any time you reset the units of measure, the test point will automati-
cally return to zero and must be reset.
8.6.3. To set the TEST POINT press the TEST POINT button. Six digits are shown on the display
and the decimal point is on the far left. Pressing the UNIT button moves the decimal point to
the right. Once the appropriate decimal place (or increment) is found, press the UP ARROW to
move the TEST POINT value to the appropriate setting. Once the test point has been reached
press the TEST POINT button again to return to normal functioning.