Torq-Tronics 2
®
/
Owners Manual / Appendix A: Serial Commands
A.5.2
Commands available only when Calibration mode is on:
To place the tester in calibration mode, enter CM <Enter> To exit calibration mode use com-
mand CMD <Enter>
Once the tester is in calibration mode, these Serial Commands are available.
A.5.2.1
Set the calibration date type: CM9mmddyyyy<Enter>
Set the calibration date, date format MMDDYYYY.
Setting Example: CM905262011<Enter>
Returns: 05262011<CR+LF>
A.5.2.2
Set the next calibration date, date format MMDDYYYY type: CM%mmddyyyy<Enter>
Setting Example: CM%905262012<Enter>
Returns: 05262012<CR+LF>
5.2.3
Set tester serial number type CM!xxxxxxxx<Enter>
Where ‘xxxxxxxx’ is the eight digit serial number of the tester. (This field is eight characters
long and must lead with a 0.)
Example: 01234567<CR+LF>
A.5.2.4
Set transducer serial number type: CMTxxxxxxxx<Enter>
Where ‘xxxxxxxx’ is the eight digit serial number of transducer. (This field is eight charac-
ters long and must lead with a 0.)
Example: 00001234<CR+LF>
A.5.2.5
Set the user string for the tester. Up to 16 characters long.
Example Type: CMPTester1234567890<Enter>
Or
Example Type: CM#Tester1234567890<Enter>
Returns: <CR+LF>
A.5.2.6
Set the user string for the transducer. Up to 16 characters long, Type: CMWTest Station
1<Enter>
Returns: <CR+LF>
A.5.2.7
Test Bi-color LEDs. Should blink: Green, then red, then yellow.
CML<Enter>
A.5.2.7.1 Set Bi-Color LEDs control for SR part number 40118 Type: CML0<Enter>
A.5.2.7.2 Set Bi-Color LEDs control for SR part number 40126 type: CML1<Enter>
A.5.3
Commands available only when Diagnostic mode is on:
(See Special ASCII command D<Enter> or D1<Enter> or D0<Enter>)
A.5.3.1 Enter Transducer Calibration Adjustment procedure Type: M<Enter>
A.5.3.2 Unlock and clear an overload transducer condition Type: Z<Enter>