Torq-Tronics 2
® /
Owners Manual
9:
How To Test a Manual Torque Screwdriver
or a Click Type Torque Wrench
9.1
Select the initial peak mode of testing by pressing PEAK
TORQUE and the UP ARROW key. Then press the PEAK
TORQUE key again.
9.2
Select the units of measure for the tool by pressing the UNITS
key and the UP ARROW until you find the unit of measure.
Then press the UNITS key again.
9.3
To set the TEST POINT press the TEST POINT button. Six
digits are shown on the display and the decimal point is on the
far left. Pressing the UNIT button moves the decimal point to the
right. Once the appropriate decimal place (or increment) is found,
press the UP ARROW to move the TEST POINT value to the
appropriate setting. Once the test point has been reached press the
TEST POINT button again to return to normal functioning.
9.4
Press the UP or DOWN ARROW to change the Percent of Accuracy Value appropriate for
the tool being tested.
9.5
When you have found the appropriate torque value range for the tool, press TEST POINT
again.
9.6
Press ZERO to clear any offset on the transducer. NEVER press ZERO to clear an offset
when a wrench is engaged in the tester.
9.7
Engage the wrench to the tester and begin testing.
9.8
The unit will autoclear after the load is removed and the test timer has expired.
10: How To Test a Beam, Dial, or Cam-Over
Torque Wrench
10.1.
Select peak mode of testing by pressing PEAK TORQUE and
the UP ARROW key. Then press the PEAK TORQUE key
again.
10.2
Select the units of measure for the tool by pressing the UNITS key
and the UP ARROW until you find your unit of measure. Then
press the UNITS key again.
10.3
To set the TEST POINT press the TEST POINT button. Six
digits are shown on the display and the decimal point is on the far left. Pressing the UNIT
button moves the decimal point to the right. Once the appropriate decimal place (or incre-
ment) is found, press the UP ARROW to move the TEST POINT value to the appropriate
setting. Once the test point has been reached press the TEST POINT button again to return
to normal functioning.
10.4
Press the UP or DOWN ARROW to change the Percent of Accuracy Value appropriate for
the tool being tested.
10.5
When you have found the appropriate torque value range for the tool, press TEST POINT
again.
10.6
Press ZERO to clear any offset on the transducer. NEVER press ZERO to clear an offset
when a wrench is engaged in the tester.
10.7
Engage the wrench to the tester and begin testing.
10.8
There is no auto-clear in track mode.
Owners Manual
Torq-Tronics 2
Sturtevant Richmont
Direct: 1 847/455-8677
U.S. Toll-Free: 800/877-1347
www.srtorque.com
All click wrenches:
use INITIAL PEAK
MODE
Torque screwdrivers:
use PEAK MODE
All beam, dial or cam-
over torque wrenches:
use PEAK MODE.
11/16/201
7
REV
E
. Document Page 15 of 39