16.4.28 Press the ‘MENU’ or ‘UP’ key to move to the next test point.
16.4.29 Again hang the weight platform (hook) and the appropriate weights (60%) (found in the
“Calibration Chart”) on the calibration arm-gently, and imparting as little motion to the
weights as possible. Wait for the arm to completely stop moving then press the ‘CLEAR’ or
‘DOWN’
key on the unit.
Observe acknowledgement of the 60% clockwise calibration point.
16.4.30 Press the ‘MENU’ or ‘UP’ key to move to the next test point.
16.4.31 Again hang the weight platform (hook) and the appropriate weights (70%) (found in the
“Calibration Chart”) on the calibration arm-gently, and imparting as little motion to the
weights as possible. Wait for the arm to completely stop moving then press the ‘CLEAR’ or
‘DOWN’
key on the unit.
Observe acknowledgement of the 70% clockwise calibration point.
16.4.32 Press the ‘MENU’ or ‘UP’ key to move to the next test point.
16.4.33 Again hang the weight platform (hook) and the appropriate weights (80%) (found in the
“Calibration Chart”) on the calibration arm-gently, and imparting as little motion to the
weights as possible. Wait for the arm to completely stop moving then press the ‘CLEAR’ or
‘DOWN’
key on the unit.
Observe acknowledgement of the 80% clockwise calibration point.
16.4.34 Press the ‘MENU’ or ‘UP’ key to move to the next test point.
16.4.35 Again hang the weight platform (hook) and the appropriate weights (90%) (found in the
“Calibration Chart”) on the calibration arm-gently, and imparting as little motion to the
weights as possible. Wait for the arm to completely stop moving then press the ‘CLEAR’ or
‘DOWN’
key on the unit.
Observe acknowledgement of the 90% clockwise calibration point.
16.4.36 Hang the calibration point (100%) weights, as found in the “Calibration Chart”, on
the arm, gently. Wait for the arm to completely stop moving then press the ‘CLEAR’ or
‘DOWN’
key on the unit. Observe acknowledgement of the 100% clockwise calibra-
tion point.
16.4.37 Press the ‘MENU’ or ‘UP’ key to move to the next test point.
16.4.38 Remove weights and reverse calibration arm to the counter-clockwise direction.
16.4.39 Hang weight platform and full-scale (100% load as found in the “Calibration Table”) load
on the arm, gently. Wait until the arm has completely stopped moving then remove the
weights and platform.
16.4.40 Wait one minute.
16.4.41 Repeat steps 38, 39 and 40 two more times, then proceed to next step.
16.4.42 Remove calibration arm from tester.
16.4.43 Press the ‘CLEAR’ or ‘DOWN’ key on the unit to capture the CCW 0 calibration point.
(Green led will flicker.)
16.4.46 Press the ‘MENU’ or ‘UP’ key to move to the next test point.
16.4.47 Again hang the weight platform (hook) and the appropriate weights (10%) (found in
the “Calibration Chart”) on the calibration arm-gently, and imparting as little motion
to the weights as possible. Wait for the arm to completely stop moving then press the
‘CLEAR’ or ‘DOWN’
key on the unit.
Observe acknowledgement of the 10% counter clockwise calibration point.
16.4.48 Press the ‘MENU’ or ‘UP’ key to move to the next test point.