Torq-Tronics 2
® /
Owners Manual
3:
Installation
3.1
Select location that provides sufficient strength and stability to resist the reaction torque that
will be generated during tester use. The location should have available the required source of
electric power and computer communications that are needed. Make sure the selected loca-
tion has sufficient room for the operator to operate the tools to be tested on the unit.
3.2
Use the Mounting Holes in the Back Plate as a template, then drill all holes as required to
solidly mount the unit to the selected location.
3.3
Use 5/16” (or M6) bolts or bolts and nuts to mount the unit in place. Grade 8 (Class 10.9)
bolts are required. Torque to 100-150 in.lb (11.3-16.9 Nm)
3.4
Slowly and carefully test the structural integrity of mount before proceeding further. If the
unit resists rotation throughout capacity, go to the next step. If the unit moves or the mount-
ing location moves, correct the problem before proceeding.
3.5
Carefully remove the Battery Compartment Cover using the handle on the cover. Install the
four (4) AA NiMH batteries into the Battery Holder. Pay careful attention to battery orien-
tation. Incorrect battery orientation will severely damage the Torq-Tronics 2 and void the
warranty. Reinstall the Battery Compartment Cover. NOTE: USING BATTERIES OTH-
ER THAN THE REQUIRED NIMH RECHARGE-ABLE BATTERIES WILL DAMAGE
THE UNIT AND VOID THE WARRANTY.
3.6
Connect the universal ROHS compliant AC/DC power supply adapter (6 volt 3 amp rated
to 100 to 200 volts 50 to 60 Hz) at the male end of the Connector that screws on tester to
ensure power tool testing does not cause the power source to come loose.
3.7
Use the RS-232 port (located near the 3 way rocker power switch) to connect to serial print-
ers or computers. A USB port adapter is supplied with Torq-Tronics 2.
3.8
Test the unit by moving the 3 way rocker power switch from off (middle position ) to either
AC power (up) or Battery power (down). When the Display comes on, the installation is
complete.
3.9
Setting the date and time.
Retrieve U.S. time date clock type: @U< Enter >
Returns format: MM/DD/YYYY,HH:MM:SS<CR+LF>
Set U.S. time date clock type: @Umm/dd/yyyy,hh:mm:ss<Enter>
mm = the two digit month range 01 – 12.
dd = the two digit day range 01-31.
yyyy = the four digit year.
hh = the two digit hour in military time range: 00 -23.
mm = The two digit minute range: 00-59.
ss = The two digit seconds range 00-59.
Returns format: @Umm/dd/yyyy,hh:mm:ss<CR+LF>
3.10
Retrieve European time date clock type: @E<CR+LF>
Returns format: DD/MM/YYYY,HH:MM:SS
Set European time date clock type: @Edd/mm/yyyy,hh:mmm:ss<Enter>
dd = the two digit day range 01-31.
mm = the two digit month range 01 – 12.
yyyy = the four digit year.
hh = the two digit hour in military time range: 00 -23.
mm = The two digit minute range: 00-59.
ss = The two digit seconds range 00-59.
Returns format: DD/MM/YYYY,HH:MM:SS<CR+LF>
Owners Manual
Torq-Tronics 2
Sturtevant Richmont
Direct: 1 847/455-8677
U.S. Toll-Free: 800/877-1347
www.srtorque.com
11/16/201
7
REV
E