15.5.7 Use appropriate calibration arm and deadweights from chart to apply torque load of
10% of full scale to the Transducer in the direction of torque to be measured. Let the
weights settle, and then record the results. Repeat this step 3 times.
15.5.8 Use appropriate calibration arm and deadweights from chart to apply torque load of
30% of full scale to the Transducer in the direction of torque to be measured. Let the
weights settle, and then record the results. Repeat this step 3 times.
15.5.9 Use appropriate calibration arm and deadweights from chart to apply torque load of
50% of full scale to the Transducer in the direction of torque to be measured. Let the
weights settle, and then record the results. Repeat this step 3 times.
15.5.10 Use appropriate calibration arm and deadweights from chart to apply torque load of
70% of full scale to the Transducer in the direction of torque to be measured. Let the
weights settle, and then record the results. Repeat this step 3 times.
15.5.11 Use appropriate calibration arm and deadweights from chart to apply torque load of
100% of full scale to the Transducer in the direction of torque to be measured. Let the
weights settle, and then record the results. Repeat this step 3 times.
15.5.12 Reverse the torque direction and repeat steps 15.5.6 through 15.5.11. When complete,
proceed to step 15.5.13.
15.5.13 Report results per your process.
15.6
Acceptance Criteria
The measured torque values from the torque test equipment must fall within required
accuracy requirements and the guard band
as listed below, at each test point within the
range specified.
ASME B107.29M
Accuracy Class
ACCURACY CLASS
% of I. V. range from 10% to 100% of Full Scale Torque Tester Model
B
+/-.5
Torq-Tronics 2