Service Information
33
SCALE SYSTEM DIAGNOSTICS AND CALIBRATION (CONTINUED)
Displaying Individual Load Cell Outputs:
A defective load cell can be detected by entering diagnostics and displaying individual load cell outputs.
1. Enter the diagnostic mode. The LCD will display “ANGLE CALIBRATE” when the diagnostic mode is
activated.
2. Repeatedly press and release the up or down arrow button (ZERO or SCALE ON/OFF) until the LCD
displays “DISPLAY CORNER CTS”.
3. Press and release the ENTER button (LBS/KGS). The LCD should display “
ƜƝ
SELECT CORNER”.
The two buttons listed below function as POSITION buttons to select the four corners of the bed’s litter.
Whenever the LCD displays “
ƜƝ
SELECT CORNER”, press one of these buttons to cycle through the
corners and to select the load cell assembly at the desired corner.
A. ZERO = cycle up through the four corners
B. SCALE ON/OFF = cycle down through the four corners
4. Press and release the position button that corresponds with the load cell to be checked. The LCD should
display “X/X=NNN.N”. ”X/X” represents the initials of the selected corner, i.e. H/R will be displayed for
the patient’s head end, right side. ”NNN.N” represents the resistance of the load cell.
5. Repeat step four for each corner. Head end weight readings will normally be lower than foot end weights.
Weight readings should be constant. A drifting 000.0 or 999.9 weight, or a reading that does not change
when weight is applied to that corner of the bed indicates a problem with the selected load cell assembly
or load cell cable.
Verifying Scale Accuracy:
1. Zero the empty bed. Place a known weight on the center of the bed; the heavier the better and no less
than 100 pounds. The displayed weight should be within
±
2% of the actual weight.
2. If the displayed weight is not accurate, remove the weight from the bed and place the scale CPU in calibra-
tion mode.
Scale Calibration:
NOTE
It requires
two people
to enable the calibration mode for the scale system.
Calibrate the scale system with a known weight of approximately 50 pounds.
1. To enter the calibration mode, unplug the bed’s power cord from the wall socket.
2. Press and
hold down
the LBS/KGS button.
3. While still holding the LBS/KGS button, plug the bed’s power cord into the wall socket.
4. After two seconds, release the LBS/KGS button. The LCD should read “ANGLE CALIBRATE”. The cal-
ibration mode is now active.
5. Press and hold the ENTER button (LBS/KGS). Zero the bed, following the displayed instructions. When
the bed is zeroed, the LCD should display “REF X100=<0>5000”. This is the factory default for 50
pounds. If 50 pounds will be used to calibrate the scale, proceed to step 7.
Summary of Contents for TriaDyne III Proventa 2030
Page 16: ...Electrical System Information 15 CPU BOARD 3002 407 950 ...
Page 32: ...Service Information 31 A B A B VIEW FROM CENTER OF BED ...
Page 48: ...Base Assembly and Options 47 HEAD END LEFT HEAD END ...
Page 49: ...Base Assembly and Options 48 Installation of Roll Pin into Brake Shaft Crank Typical 4 Places ...
Page 50: ...Base Assembly and Options 49 HEAD END HEAD END ...
Page 51: ...Base Assembly and Options 50 HEAD END LEFT SIDE FOOT END ...
Page 52: ...Base Assembly and Options 51 Steer Option ...
Page 53: ...Base Assembly and Options 52 Typical Both Ends Foot End Head End ...
Page 56: ...Lift Assembly Common 55 Assembly part number 3001 245 275 reference only ...
Page 67: ...Litter Assembly and Options 66 ...
Page 68: ...Litter Assembly and Options 67 ...
Page 69: ...Litter Assembly and Options 68 ...
Page 70: ...Litter Assembly and Options 69 ...
Page 71: ...Litter Assembly and Options 70 HEAD WALL VIEW FOOT SECTION ...
Page 72: ...Litter Assembly and Options 71 CPR RELEASE ...
Page 73: ...Litter Assembly and Options 72 CPR SLIDE BRACKET DETAIL SMART TV OPTION ...
Page 74: ...Litter Assembly and Options 73 GROUNDING DETAIL HEAD END FOOT END ...
Page 75: ...Litter Assembly and Options 74 HEAD END FOOT END SIDERAIL EXTENSION CABLE ROUTING ...
Page 82: ...Head End Siderail Assembly 81 LEFT SIDERAIL SHOWN ...
Page 83: ...Head End Siderail Assembly 82 CARRIER ARM ASSEMBLY ...
Page 84: ...Head End Siderail Assembly 83 INNER SIDERAIL PANEL OUTER SIDERAIL PANEL ...
Page 85: ...Head End Siderail Assembly 84 LEFT HEAD END SIDERAIL WIRE ROUTING ...
Page 86: ...Head End Siderail Assembly 85 RIGHT HEAD END SIDERAIL WIRE ROUTING ...
Page 96: ...Foot End Siderail Assembly 95 Outer Siderail Panel Outer Siderail Panel Left Side Shown ...
Page 97: ...Foot End Siderail Assembly 96 Left Side Shown ...
Page 98: ...Foot End Siderail Assembly 97 Outer Siderail Panel Inner Siderail Panel ...
Page 99: ...Foot End Siderail Assembly 98 Foot End Siderail Wire Routing Left Side Shown ...