39
S
Figure 11.12- - Basic Network Diagram
10.3 Site Preparation Requirements
The following provide instructions that must be completed before installation.
10.3.1 SuiteStatus System
Each SuiteStatus System can support up to sixty-four (64) cameras and three (3) All-In-One Viewing
Systems. Each All-In-One Viewing System and SuiteStatus camera must reside on the same network
with one CAT-5e (or above) Ethernet cable connection at each device.
10.3.2 Power
The All-In-One Viewing System(s) must have one (1) standard electrical outlet available. Each Suit-
eStatus camera must have one (1) standard electrical outlet (12V DC outlet) or Power-over-Ethernet
(PoE) connection that is located where the camera will reside.
10.3.3 Mounts
A wall or ceiling mount is required when installing a camera; therefore, the number of mounts must
match the total number of cameras. Please consult with the Project Manager if there is a discrepancy.
The ceiling mount will require a 184mm diameter circle in the ceiling where the camera will reside.
The wall mount will require 3, 5/16” holes to be drilled into the wall. Plastic dowels will be included
with each camera to fill these holes. For both mounts, ensure that an Ethernet cable extends to the
mount to supply the camera with network and/or power connectivity.
10.3.4 External Display
External displays are available for purchase to duplicate the screen of the All-In-One Viewing System.
Consult with the Project Manager and/or Engineer to ensure the correct parts are on order.