15
S
Fi
gu
re 5.5 - Th
re
e-B
ay D
ocu
m
en
ta
tio
n S
ta
tio
n
4"(80mm) x 4"(80mm) ELECTRICAL J-BOX
4"(80mm) x 4"(80mm) ELECTRICAL J-BOX
FLEX CONDUIT
WALL
ELECTRICAL DUPLEX
(6X) ELECTRICAL
QUAD
(2X) ELECTRICAL DUPLEX
CUT-OUT
CUT-OUT
DATA / AV
CONNECTION
FLEX CONDUIT
4"(80mm) x 4"(80mm) ELECTRICAL J-BOX
4"(80mm) x 4"(80mm)
ELECTRICAL
J-BOX
18" x 18" SWITCHPOINT INFINITY J-BOX
6"(150mm) X 6"(150mm) J-BOX (FOR STRYKER
LIGHTS ONLY)
(6X) ELECTRICAL
QUAD
15"
20"
6"
1.
THIS DRAWING IS FOR REFERENCE OF CUT-OUT LOCATIONS ONLY. IT IS NOT INTENDED FOR MANUFACTURING USE.
2.
DRAWING DISPLAYS FRONT VIEW OF DOCUMENTATION STATION WITH DOORS OMITTED.
3.
PROVIDE AND INSTALL SIX (6) ELECTRICAL QUAD OUTLETS AND SIX (6) DUAL GANG FACE PLATES AS WELL AS TWO (2) ELECTRICAL DUPLEX OUTLETS WITH TWO (2) SINGLE GANG FACE PLATES INTO THE ELECTRICAL JUNCTION BOXES PROVIDED.
4.
INSTALL AND CONNECT FLUORESCENT LIGHTING FIXTURE IF PROVIDED.
5.
PROVIDE AND CONNECT 20 AMP CIRCUITS FOR THE STRYKER DOCUMENTATION STATION FOR EACH QUAD PLEX OUTLET: ONE (1) 20 AMP CIRCUIT FOR EACH OF THE TWO (2) DUPLEX OUTLETS IN THE DOCUMENTATION STATION WITH ONE (1) CIRCUIT ALSO POWERING THE FLUORESCENT LIGHTING. IF INSTALLING STRYKER VISUM LIGHTS, ONE (1) 20 AMP CIRCUIT IS REQUIRED FOR EACH LIGHT POWER SUPPLY BOX (REFER TO VISUM PRE-INSTALLATION GUIDE FOR LOCATION OF POWER SUPPLY BOX). POWER SHOULD BE CONNECTED USING CONDUIT FROM THE WALL TO FACILITATE FUTURE ACCESS TO THE CABLE CHASE IN THE BACK OF THE DOCUMENTATION STATION. ALL STRYKER DOCUMENTATION STATION CIRCUITS SHOULD BE ON CRITICAL POWER.
SCALE: 1" = 1'
6X 4.25 X 4.25 CUTOUT (ELECTRICAL QUAD)
6X 2.5 DIA. (GROMMET)
14.0 X 20.0 CUTOUT
2.25 X 4.13 CUTOUT
(DATA / AV)
CUTOUT
10.0"
11.2"
21.7"
23.8"
16.2"
4.8"
7.4"
5.1"
3X 66.0"
78.5"
29.1"
41.8"
18.0"
7.8"
14.0"
21.9"
28.0"
DOC STATION CENTER POINT
14.0 X 14.0 CUTOUT (BEHIND SHELVES) IF HARD-PIPED
90.0"
www.stryker.com
A
FLOWER MOUND, TEXAS 75028
1410 LAKESIDE PARKWAY, SUITE 100
PHONE: (877) 789-8106
FAX: (972) 410-7001
SHEET:
B
C
16.3"
28.5"
1.4"
NOTES: (UNLESS OTHERWISE SPECIFIED)
1. EQUIPMENT LIST:
02
02