26
S
7. Level Systems Room
Note
All cabling must be provided and pulled by Hospital/Contractor prior to Stryker's
arrival.
Level System's are most often installed in integrated doctor's offices, small conference rooms, and
pathology or nuclear medicine labs. This control system is placed in a cabinet next to or in the room's
cabinetry or desk. Cable passage must be allowed to the junction box for all equipment placed on the
desk or in the cabinetry.
7.1 Customer Supplied Casework (CSC)
Data
Per listed equipment
Back Box
Per listed equipment
Power
One (1) - 20 amo circuit for outlets behind video router
Space
Requirements
• Must allow for a minimum 2" cable passage between all components housed
inside.
• Section housing video router must have an interior dimension of at least 21"W
X 16"H X 22"D
• Section housing router must be vented.
• Must allow for direct access to back boxes per requirements listed below.
Note
Customer/Contractor responsible for installation of casework prior to Stryker
installation.
7.2
Casework Equipment
7.2.1
Level One System (S1)
LA
Dimensions
20.5"W X 8.5"H X 16"D
Data
None
Back Box
• One (1) 6"W X 6"H X 4"D (minimum) junction box flush mounted at
same height as Level System
• Terminate all integration conduits to this junction box
Power
One (1) duplex outlet next to junction box
7.2.2
Level Two System (S2)
LA
Dimensions
20.5"W X 14"H X 19"D
Data
None
Back Box
• One(1) 12"W X 12"H X 4"D (minimum) junction box flush mounted at
same height as Level System
• Terminate all integration conduits to this junction box
Power
One (1) duplex outlet next to junction box