![Stryker Operating Room Information System Preinstallation Manual Download Page 23](http://html.mh-extra.com/html/stryker/operating-room-information-system/operating-room-information-system_preinstallation-manual_1385978023.webp)
19
S
VISUM LED POWER SUPPLY BOX (N1)
Dimensions
• PSB: 12.5"W X 7"H X 15.5"D
• Total Space Required: 20"W X 9"H X 21"D
Back Box
• One (1) 6"W x 6"H x 4"D (or larger) junction box flush mounted.
• Set bottom of box 14" above finished floor.
• Terminate all surgical light conduits to this junction box
5.4
iSuite Equipment Integration Notes
Note
All conduit runs include insulated bushings and pull strings.
Note
Conduit runs cannot exceed 50' from end-to-end. Do not exceed four (4) 90 de-
gree bends.
5.4.1
Wall Mounted 46" or 55" LCD Monitor or 42" TouchPanel
Conduit
One (1) 1 1/4" conduit
Back Box
• 4"W X 4"H junction box with single-gang mud ring
• Mounted directly above the top of the mounting bracket.
Power
One (1) standard duplex outlet mounted adjacent to junction box
Structural
• Customer/Contractor to mount Stryker provided bracket to the wall in the de-
sired location with proper reinforcement to support the monitor prior to Stryker
installation.
• Stryker Project Manager will provide mounting specifications.
ITEM-ITEM
QTY
SIZE
WMM – A
1
1 1/4"
5.4.2
Wall/Ceiling Mounted Room Status Camera (WRC, CRC)
LA
Conduit
One (1) ¾" conduit terminated at nearest corridor cable tray
Back Box
• 4"W x 4"H junction box with single-gang mud ring
• Flush mounted at 12" below finished ceiling (wall mounted only)
Power
None
Cabling
• One (1) power cable and one (1) video cable
• One (1) Belden 8723 / 88723 (maximum 1000')
• One (1) Belden 8241 / 88241 (maximum 1000')
• Cables require 15' service loop at both ends
Note
All Status System cabling must be provided and pulled by the Hospital/Contrac-
tor prior to Stryker Installation.