![Stryker LUCAS 3 Service Manual Download Page 70](http://html.mh-extra.com/html/stryker/lucas-3/lucas-3_service-manual_1385958070.webp)
LUCAS
®
3 Chest Compression System –
Service Manual
3328798-002, ©2017 Physio-Control, Inc.
Page
70
of
91
2. Reassembling
•
Put back the Communication PCBA according to
Reassembling
instruction in
•
If the SD-card from the old PCBA is working properly insert the old SD-card into the new PCBA, otherwise
keep new SD-card installed.
•
Put the new Controller PCBA in place and fasten loosely with two screws together with the Bottom Plate
with a
Torx 25 (T25)
Screwdriver.
•
Put back the Compression Module according to
Reassembling
•
Put back the Hood according to
Reassembling
instruction in
•
Put back the Bellows according to
Reassembling
•
Put back the Suction Cup.
3. Programming
•
Connect a laptop and program the Controller PCBA according to
•
Check that “Device S/N”, Time and Date, “Total Number of Strokes”, “Hardware ID” and “Device Friendly
Name” are correct (if possible check against “Store View” performed in Disassembling step). If “Store View”
was not possible set: “Device S/N” (e.g. 35160021) and “Hardware ID” (e.g. 100921-00) according to Type
Label, “Total Number of Strokes” to 0 (zero) and “Device Friendly Name” to “LUCAS3-“ + serial number
according to Type Label.
4. Test
Perform a Function Check according to
How to Replace the Protective PCBA
1. Disassembling
•
Before disassembling and changing any PCBA and if possible; connect a laptop and perform “Store View”
using
•
Follow
Disassembling
instructions in
How to Replace the Controller PCB
to take out the Protective PCBA.
•
Warning
: Always use ESD protection when handling PCBAs.
2. Reassembling
•
Put the new PCBA in place and fasten loosely with two screws together with the Bottom Plate with a
Torx
25 (T25)
Screwdriver
.
•
Put back the Compression Module according to
Reassembling