LUCAS
®
3 Chest Compression System –
Service Manual
3328798-002, ©2017 Physio-Control, Inc.
Page
50
of
91
T1 BATT CS
ALARM
Battery
communication
check- sum fault
Connector fault.
Check
connections.
Unplug and insert
battery, check for
fault. Or replace
battery.
None.
Shall be shown at
LOG file read.
Battery not
recognized as
Physio-Control
proprietary (no
S/N)
T1 BATT
COPYRIGHT
ALARM
Incorrect
copyright string
read from
battery
Incorrect battery
type
or communication
fault
Check
connections.
Unplug and insert
battery, check for
fault. Or replace
battery.
None.
Shall be shown at
LOG file read.
Battery not
recognized as
Physio-Control
proprietary (no
S/N)
TOO MANY
USES
Inserted battery
is used more
than 200 times.
Worn out battery.
Battery
Orange battery LED
(B1)
Battery
replacement
recommended
(S/N xxxxxxxx)
TOO OLD
BATTERY
Inserted battery
is too old.
Too old battery
used.
Battery
Orange battery LED
(B1)
Battery
replacement
recommended
(S/N xxxxxxxx)
User Errors
Indication
Probable Cause
Probable
Consequence
User Observation
LRG Text
The new Start
position was not
accepted.
Too small patient Start.
Checked by POT graph
data > 133.
Too small patient, use
manual compressions.
Not accepted start,
information signal
(3 fast signals).
Too small
patient.
Too fast mode change.
Repeated press on
button.
Not accepted start,
Cup moved/held
down at PAUSE
Cup moved down 10mm
by user at PAUSE or Too
high start position result
in 10mm 13N spring
release at next PAUSE
(after Active).
Too high Start position
resulting too shallow
compressions.
13N release
information signal (3
fast signals).
Start Position
adjustment
required
Stop/Adjust direct to
Active (RUN/30:2)
The user wants to start
quickly.
-
Quick start.
Max Quick Fit
distance done.
Too high placed Cup
above patient chest prior
to start.
Too high Start position
resulting too shallow
compressions.
Free distance
between chest and
Cup.
High Preload detected
at Start position.
User pressed down the
Cup during Start position
set. Patient moved.
High down force at
zero position.
Compressed chest.
Audible Timer Alert.
Timer time out
Timer
Audible alert
Timer Alert.
Rate is changed.
ACTIVE mode and used
ACTIVE button pressed
to change Rate.
Compression
frequency changed
Rate change
Rate change to
xxx
Bluetooth Connection Error
When connecting LUCAS 3 via Bluetooth to
DT Express, LUCAS Report Generator
or
LUCAS Program Loader
, if
connection errors happened repeatedly, the possible cause might be the Bluetooth driver on computer is not up to date.