![Stryker LUCAS 3 Service Manual Download Page 65](http://html.mh-extra.com/html/stryker/lucas-3/lucas-3_service-manual_1385958065.webp)
LUCAS
®
3 Chest Compression System –
Service Manual
3328798-002, ©2017 Physio-Control, Inc.
Page
65
of
91
•
Carefully spread the PCB boards and lift out the Copression Module, make sure that no cables get stuck or
harmed especially the Linear sensor Cables.
2. Reassembling
•
Carefully lower the Compression Module down between the PCB’s.
•
Position the PCB boards and fasten the Compression Module in the PCB brackets, use thread lock fluid
and tighten to 9.0 in-lbs / 1.0 Nm with a Torque Screwdriver
Torx 10 (T10).
•
Put back the Fan Bracket, use thread lock fluid and tighten to 35 in-lbs / 4.0 Nm with a Torque
Screwdriver
Torx 25 (T25).
•
Fasten the Compression Module with four screws, use thread lock fluid and tighten to 35 in- lbs / 4.0 Nm
with a Torque Screwdriver
Torx 25 (T25).
•
Unscrew the two screws on the bracket that are loose, apply thread lock fluid and tighten to 35 in-lbs / 4.0
Nm with a Torque Screwdriver
Torx 25 (T25).
•
Re-connect all cables according to picture in
Disassembling
.
•
Replace removed Cable Ties (if applicable).
•
Put back the Hood according to
Reassembling
instruction in
•
Put back the Bellows according to
Reassembling
•
Put back the Suction Cup.
3. Test