LUCAS
®
3 Chest Compression System –
Service Manual
3328798-002, ©2017 Physio-Control, Inc.
Page
86
of
91
Replacing the Main Body/Frame Assembly
4.
Disassembling
•
Follow the instructions in Section Replace the Compression Module.
•
Follow the instructions in Section Replace the Controller PCB
•
Follow the instructions in Section Replace the Protective PCB
•
Follow the instructions in Section Replacing the Power Inlet.
•
If replacing the Main Body follow the instructions in Section Replace the Support Legs.
Warning
: Always use ESD protection when handling PCB’s!
5.
Reassembling
•
Re-install the Power Inlet using instructions in section Replacing the Power Inlet.
•
Install the Main Body Lid and tighten all four Torx 20 screws to 13 in-lbs / 1.5 Nm.
•
Re-install the Compression Module, Controller PCB, Protective PCB, and Hood.
•
If removed, re-install the Support Legs.
6.
Test
•
Perform a Function Check according to Section Function Check.