Remove the Bellows by treading it over the Piston.
2. Reassembling
•
Carefully thread the Bellows over the piston and position it as far up as possible.
•
Place the clamp and fixate the bellows by tightening the two screws to 9.0 in-lbs / 1.0 NM with a Torque
Screwdriver
Torx 10 (T10) /
or 13.0 in-lbs / 1.5 NM with a
Torx 20 (T20)
depending on Bellows Clamp
.
The clamp should be positioned as far up as possible.