Place the Bellows ring above the retaining ring.
•
Thread the edge of the Bellows over the Bellows ring.
•
Put back the Suction Cup.
3. Test
Perform a Function Check according to
How to Replace the Compression Module
1. Disassembling
•
Take off the Suction Cup and remove the Bellows according to
Disassembling
instruction in
•
Remove the Hood according to instructions in
•
Disconnect the following cables according to pictures below
1.
Protective Linear sensor Cable (on the
back side of Protective PCB)
2.
Communication Cable
3.
Motor Rotation Sensor Cable
4.
Control Linear sensor Cable
5.
Speaker Cable
6.
Fan Cable
7.
Antenna Cable