View Error Code
Once the connection is made from Bluetooth or USB cable, the information about the device should appear in the
designated CPU windows.
The device Errors and Warnings will appear in each CPU’s box window.
To view existing log information on the device, click
LOG
button,
, Log view shows log data on screen.
Highlight a row
•
Press
READ
to view log data
•
Press
USE
to see the Error and Warnings code in different CPU windows
•
Press
SAVE
to save the raw log file data to PC (only for developers to use)
•
Press
EXIT
to exit LOG view
Example:
1. Highlight a row according to
Date
and
Time
the error occurred from LOG view;
2. Click
READ
and wait to see the “
Collect errors and warnings from LOG File
” to show up on the left side of
window;
3. If the
Number of Errors
or
Warnings
are not 0, Click
USE
to see the highlighted LOG Error code in the CPU
windows (
USE
will exit LOG view also).
- Picture shows LOG File content on the left for LUCAS 3 Version 3.0 (Part No: 250039-00)
Note, differences in J-numbers and CheckSum Numbers may occur due to different program versions –
3.
2.
1.