LUCAS
®
3 Chest Compression System –
Service Manual
3328798-002, ©2017 Physio-Control, Inc.
Page
19
of
91
Connect the device via USB cable
1. Remove the hood according to
Disassembling
, and use the Hood
Holder bracket to hold the hood on the side and the battery should be installed.
2. Connect an USB cable between the computer and the device, the connector is situated at the lower left side of
the Control PCB.
3. Start the device
4. Start the software by executing LUCAS3_V3_0.exe file,
.
(Note. Different software versions might occur)
5. Normally the software detects the device at start. If not, then Click
Init USB
.
6. When USB connection is established, the Program Loader screen will show CPUs information
Hood Holder Bracket
USB Port Location
Hood Holder Bracket Location