Lock the four screws with thread lock fluid and tighten to 5.3 in-lbs / 0.6 Nm with a Torque Screwdriver
PH1
.
3. Test
Perform a Function Check according to
How to Replace the Bellows
1. Disassembling
•
Take off the suction cup
•
Remove the clamp by removing the two screws with a
Torx 10
(
T10
) or
Torx 20 (T20)
Screwdriver.
•
Remove the Bellows ring, start by treading the bellows over the ring, then continue to take off the ring as
showed in the picture.
Correct hood installation