EN-5
4. When the Crossfire 2 system and physiological monitoring equipment are used
simultaneously on a patient, position any monitoring electrodes as far as possible from the
surgical electrodes. Monitoring equipment using high frequency, current-limiting devices is
recommended. Needle monitoring electrodes are not recommended.
5. During use, operators should wear standard surgical gloves to help reduce the risk of electric
shock.
6. To prevent patient injury, select the lowest output power required for the intended purpose.
7. Do not exceed the rated accessory voltage of electrosurgical accessories. Only use
electrosurgical accessories that have a rated accessory voltage equal to or greater than the
maximum output voltage of the generator.
8. Do not activate the Crossfire 2 system until the probe is properly positioned in the patient.
9. Ensure that the probe tip, including the return electrode, is completely surrounded by
irrigant solution during use.
10. Maintain the active electrode in the field of view at all times to avoid tissue damage.
11. Keep active electrodes isolated from the patient when not in use.
12. When not in use, remove the handpiece and disposable attachments from the surgical
site and place them away from metallic objects. Attachments should be separated from
other electrosurgical equipment to avoid inadvertent electrical coupling between devices.
Inadvertent activation may cause user/patient injury and/or product damage.
13. Failure of the system may result in an unintended increase in output power.
14. Neuromuscular stimulation may occur when RF probes are used.
15. Smoke generated during electrosurgical procedures may be harmful to surgical personnel.
Take appropriate precautions by wearing surgical masks or other means of protection.
Cautions
To avoid product damage, observe the following cautions.
1. While using the handpiece, do not touch the attachment to metal objects, such as an
endoscope or metal cannula. Damage to the attachments or other devices may result.
2. Attempt no internal repairs or adjustments, unless specified otherwise in this manual. Units
requiring repair should be returned to Stryker.
3. Pay close attention to the care and cleaning instructions in this manual. Failure to follow
these instructions may result in product damage.
4. Do not remove the cover of the console as this could cause electric shock and product
damage.
Summary of Contents for Crossfire 2
Page 1: ...Crossfire 2 Integrated Resection and Energy System 0475 100 000 ...
Page 2: ......
Page 4: ......
Page 6: ......
Page 44: ......
Page 88: ......
Page 130: ......
Page 174: ......
Page 218: ......
Page 262: ......
Page 264: ......
Page 308: ......
Page 350: ......
Page 390: ......
Page 430: ......
Page 470: ......
Page 514: ......
Page 558: ......
Page 596: ......
Page 644: ......
Page 654: ...CHS 650 2 连接交流电源 3 连接机头和脚踏开关 注 释 如果连接了 过期或已使用过的 附 件 控制台 会显示一条错误消息 4 连接抽吸管 用于所有可抽吸设备 ...
Page 682: ......
Page 720: ...JA 716 米国製 注意 アメリカ合衆国連 邦法に基づき 本装置は 医師による指示または注 文に応じてのみ販売され ます 壊れ物注意 使用方法を参照してくだ さい ...
Page 722: ......
Page 761: ......