![Stryker 6086 Operation & Maintenance Manual Download Page 18](http://html.mh-extra.com/html/stryker/6086/6086_operation-and-maintenance-manual_1385832018.webp)
18
6086-009-001 REV B
www.stryker.com
Cot Fastener Installation
WALL OF VEHICLE
ANTLER FLOOR
PLATES
14 5/8” (REF.)
OR DESIRED
POSITION FROM
WALL
REAR OF
AMBULANCE
(REF.)
COT CENTER LINE
COT PERIMETER
COT
HEAD
END
12 5/8”
2 5/8”
37 5/8”
14 15/16”
6 1/8”
24”
(REF.)
80” (REF.)
Figure 4: Installation Specifications - Floor Mount Fastener
Figure 5: Installation Specifications - Wall Mount Fastener
RAIL SUPPORT BRACKET
FLOOR PLATE
WALL OF VEHICLE
ANTLER FLOOR PLATES
14 5/8”
REAR OF
AMBULANCE
(REF.)
COT CENTER LINE
COT PERIMETER
COT
HEAD
END
14 15/16”
6 1/8”
24” (REF.)
80” (REF.)
1 1/8”
3 5/8”
4 5/16”
6 1/2”
37 3/8”
WALL MOUNTING BRACKET
FLOOR OF
VEHICLE
(32 cm)
(6,6 cm)
(95,6 cm)
(37,9 cm)
(15,6 cm)
(61 cm)
(37,1 cm)
(37,1 cm)
(9,2 cm)
(94,9 cm)
(203 cm)
(16,5 cm)
(10,9 cm)
(37,9 cm)
(15,6 cm)
(2,9 cm)
(61 cm)
(203 cm)
Note:
These installation instructions
are intended for cots that you will NOT use with Power-LOAD. For Model 6086
cots with the Power-LOAD option, see the Power-LOAD Operations/Maintenance Manual (6390-009-001) for installation
instructions.
Summary of Contents for 6086
Page 2: ......
Page 75: ...Return To Table of Contents www stryker com 6086 009 001 REV A 75 Cot Assembly ...
Page 76: ...Return To Table of Contents 76 6086 009 001 REV A www stryker com Cot Assembly ...
Page 79: ...Return To Table of Contents www stryker com 6086 009 001 REV A 79 Base Assembly ...
Page 132: ...Return To Table of Contents 132 6086 009 001 REV A www stryker com Optional Gatch Assembly ...
Page 157: ......