Recommended safety distances between portable and
mobile HF communications equipment and the D-ACTOR
®
200
The D-ACTOR
®
200 is intended for use in an electromagnetic environment in which radiated HF
disturbances are controlled. The customer or the user of the D-ACTOR
®
200 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile HF
communications equipment (transmitters) and the D-ACTOR
®
200 as recommended below, according to
the maximum output power of the communications equipment.
Rated power
of transmitter
[W]
Safety distance according to frequency of transmitter [m]
150 kHz to 80 MHz
d = 1.2
√
P
80 MHz to 800 MHz
d = 1.2
√
P
800 MHz to 2.5 GHz
d = 2.3
√
P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended safety distance can
be estimated using the equation applicable to the frequency of the transmitter, where P is the rated power
of the transmitter in watts [W] according to the transmitter manufacturer.
NOTE 1
An additional factor of 10/3 was used for calculating the recommended safety distance of transmitters in
the frequency range from 80 MHz to 2.5 GHz in order to reduce the probability that a mobile/portable
communications device brought into the patient area might inadvertently lead to a malfunction.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.