19 620 02 0117
53
Operation
4.7
Start-up
• Set the energy of the pulses to an initial value of 2 bar
eff
.
The maximum pressure is limited to 5.0 bar
eff
. The minimum pressure that can be set
is 0.3 bar
eff
.
The D-ACTOR handpiece can be operated in single pulse mode and in continuous
pulse mode.
• To work in D-ACTOR single pulse mode, select the ‘-’ symbol (dash) in the
‘Frequency’ selection box and activate the trigger button.
• To work in D-ACTOR continuous pulse mode, select a continuous pulse frequency
in the range from 1,0 to 21 Hz in the ‘Frequency’ selection box.
• Activate the trigger button.
NOTE
If the set nominal pulse value (e.g. 400 pulses) is reached during treatment the
handpiece automatically stops releasing pulses. Further treatment is possible.
As soon as a multiple of the set nominal value is reached (e.g. 800 pulses, 1,200
pulses, etc.) the handpiece stops anew.
4.8
Functional checks
Perform the following functional checks after the system has been installed:
• Check the device and the handpieces for any signs of damage.
• Put the device into operation.
• Set the energy level to 2 bar
eff
.
• Reset the treatment pulse counter on the handpiece display.
• Release individual pulses in single pulse mode.
• Release pulses in continuous pulse mode (pulse frequency 5 Hz/15 Hz).
• Press foot switch while the VACU-ACTOR is connected if it is used one.
• Check that the triggered pulses are correctly counted on the treatment pulse counter.