Operation
4.9
Standard settings
• Before each treatment, make sure that the pulse counter is set to zero.
NOTE
Set the nominal value counter to the required value.
D-ACTOR
• Start the D-ACTOR treatment at a pressure of 2 bar
eff
and a frequency of 5 Hz.
V-ACTOR
• Without Tablet:
The treatment parameters for the V-ACTOR ll are fixed at an energy level of
2,4 bar and a frequency of 31 Hz
.
With Tablet:
The treatment parameter are adjustable in an energy level range from 1-5 bar and
frequency range from 1-35 Hz.
4.10
Treatment with D-ACTOR or V-ACTOR handpiece
Safety information
Before using the device, the user must make sure it is functioning safely and in proper
condition.
• Read
C
hapter
before beginning treatment.
CAUTION!
Handpiece not positioned correctly.
Impairment to health due to ineffective treatment!
•
Define the treatment zone and make sure that the handpiece
position always corresponds to the treatment zone.
• Make sure that the treatment is only administered by users who
meet the conditions in
C
hapter
•
For safety reasons, using the device for applications other than those specified in
C
hapter
is not permitted!