19 620 02 0117
57
Operation
4.11
Treatment with VACU-ACTOR
CAUTION !
Do not apply the VACU-ACTOR in the face.
Do not apply the VACU-ACTOR longer than 30 minutes.
Do not use damaged suction cups.
Only perform treatment in a dry state.
Blood cupping is not allowed!
Only perform treatment at healthy and intact skin!
Avoid hematoma, pain and tissue damage!
• Connect the suction cup into the coupling of the air hose (see Fig. 3-11).
NOTE
The maximum underpressure of the compressor is -935 mbar.
While treatment pressures between -500 mbar and +200 mbar will be reached
depending on the adjusted intensity.
The vent hole limits the maximum underpressure to -500 mbar.
• In case of an error the vent hole can be opened.
Treatment with VACU-ACTOR can either be activated and deactivated with the
footswitch or optional can be switched on and off with the aid of the Tablet PC.
Using the Tablet PC the treatment parameters can be adjusted.
4.11.1 Using the foot switch
When using the foot switch to activate the VACU-ACTOR the treatment is carried out
with intensity 3.
The treatment will always be carried out with mode VI in which alternately an under-
pressure and overpressure will be produced.
• Apply massage oil onto the area to be treated.
• Place the VACU-ACTOR air tight to the skin.
•
Keep the vent hole closed with your finger.
• Pressing the foot switch
-->
treatment will be started.
• Pressing the foot switch a second time
-->
treatment will be stopped.
NOTE
Move the suction cup slowly over the area to be treated as long as the treat-
ment lasts.