19 620 02 0117
47
Operation
4.5.13 Exporting treatment data
This function can be used to export treatment data to a USB memory stick as
Excel-readable files.
• Ensure that your USB stick supports the USB V1.1 protocol. You can order a
validated USB stick from your dealer.
•
Load the parameter record for a specific patient.
• Press
.
– You will be prompted to connect a USB stick.
•
Confirm by pressing
.
The data is transferred once the USB connection has been established. The export file
name of the patient record is
protocol_name.csv
.
• Wait until the ‘Export completed’ message appears on the display and then
remove the memory stick.
4.5.14 Deleting patient datasets
• Open the patient record to be deleted.
• Press
.
– The dataset is set to editing mode.
• Press
.
•
Confirm your input by pressing
.
h
The patient dataset is deleted.
4.5.15 Resetting the treatment pulse counter
• Press the
RESET
button in the
COUNTER READINGS
field to set the
display of applied pulses back to ‘0’.
Fig. 4-36 Counter readings
h
The display of the number of pulses and the energy output per treatment is reset.