2-13
Uncrating Instructions
Uncrating/Installation Instructions
P-920010-163
Service Access Panels
12345678901234567890123456789012123456
12345678901234567890123456789012123456
12345678901234567890123456789012123456
12345678901234567890123456789012123456
To access the chemical injection pumps, first remove the service
access panels.
Reliance 200 Glassware Washer (see Figure 2-9.)
1.
Open detergent compartment door.
2.
Remove flat head screw holding service access panel to frame.
3.
Remove service access panel.
Reliance 250 Glassware Washer (see Figure 2-10.)
1.
Pull out service access panel.
2.
Remove two bolts securing protection plate to frame.
3.
Pull out protection plate.