4-1
Installation Checklist
Uncrating/Installation Instructions
P-920010-163
INSTALLATION CHECKLIST
4
Installation Checklist must be completed after the equipment is
installed and prior to performing the operational test (Section 5) to
ensure complete and correct installation.
NOTE: Contact your STERIS representative to schedule a techni-
cian to test your installation and demonstrate proper equipment
operation.
❏
Shutoff valves (not by STERIS), for maintenance purposes and
capable of being closed, installed on water lines and in compli-
ance with local occupational health and safety regulations, as well
as electric and plumbing codes for any special requirements that
may pertain to installation of this unit.
❏
Disconnect switches (not by STERIS), capable of being locked in
OFF
position only, installed in electrical supply line near the unit
and in compliance with local occupational health and safety
regulations, as well as electric and plumbing codes for any
special requirements that may pertain to installation of this unit.
NOTE: If unit is installed next to other equipment, shutoff valves and
disconnect switches should be located so that service can be shut
off to one piece of equipment at a time.
❏
Unit is positioned, as shown on Equipment Drawing, with required
clearance space and in relation to building supply lines.
❏
Unit is level.
❏
Building cold water line supplies water to unit as specified on
Equipment Drawing.
❏
Building hot water line supplies water to unit temperature as
specified on Equipment Drawing.
❏
Pure water line supplies water to unit as specified on Equipment
Drawing (if equipped with Pure Water Rinse option).
❏
If required, verify all water pressure regulators are installed as per
instructions provided with the installation kit, and that pressure is
adjusted as specified on Equipment Drawing.
❏
Building waste line is connected to unit as specified on Equip-
ment Drawing.
❏
Electrical supply for unit is as specified on Equipment Drawing.
❏
Floor surrounding unit has nonslip and noncombustible surface.
❏
Wall behind unit has a noncombustible surface.