1-2
P-920010-163
Uncrating/Installation Instructions
Listing of Warnings, Cautions and Symbols
WARNING – CHEMICAL BURN/EYE INJURY HAZARD:
Wear gloves and eye protection when using a chemical product. Avoid contact with eyes or skin. If
spilled or splashed, flush with water for 15 minutes. If swallowed, DO NOT induce vomiting.
Administer an alkali with plenty of water. Seek medical attention immediately.
Detergents are caustic and can cause adverse effects to exposed tissues. Do not get in eyes, on skin,
or attempt to ingest by mouth.
• Read and follow the precautions and instructions on the detergent label and in the Material Safety
Data Sheet (MSDS) prior to handling the detergent, containers or servicing the detergent injection
pumps.
• Wear protective clothing, gloves, and face shield whenever handling detergents or servicing the
detergent injection pumps and lines.
CAUTION - POSSIBLE EQUIPMENT DAMAGE:
When removing adhesives from stainless steel, use a small amount of non-flammable cleaning
solvent. Rub in a back-and-forth motion (in same direction as surface grain). Solvent rubbed in a
circular motion or applied with a wire brush or steel wool on door and chamber assemblies can be
harmful to stainless steel. Do not use solvents on painted surfaces.
When unit is installed on a combustible floor and/or close to a combustible wall, the floor and/or wall
must be covered with a metal sheet extending to the outer edges of the machine.
Before operating equipment, always firmly push the manifolded accessory rack into the manifold
connector at the back of the wash chamber. If rack is not positioned correctly, door will not close,
damage may result, and unit will be unable to effectively wash load.
Always use a silicone lubricant to lubricate squeeze tubes. Petroleum-based lubricants, such as
Vaseline or grease, will cause squeeze tubes to melt.
Do not cut flexible hose (utilities and draining). The whole 105" (2,667 mm) length of the flexible hose
is required when pulling out the unit for service or maintenance purposes.
Do not cut main electric supply cord. The whole 105"(2,667 mm) length of the main supply cord is
required when pulling out the unit for service or maintenance purposes.
Do not cut printer cable. The whole 105" (2,667 mm) length of the printer cable is required when
pulling out the unit for service or maintenance purposes.
On junction box, secure main supply cord with a strain relief to prevent damage to wiring.