2-3
Uncrating/Installation Instructions
Operator Manual
122996-004
1. Review the installation requirements:
a. Clearance - The clearance space shown on the equipment drawing
indicates the space necessary for easy installation and proper operation
and maintenance of the Basil
®
1100 Bottle Filler.
b. Utility service lines:
• Disconnect switch (not by STERIS) must be installed in the electric
supply line near the bottle filler (see Figure 2-2). Disconnect switch
must be capable of being locked in the OFF position only.
• If this machine is installed next to other equipment, disconnect switch
should be located so that service can be shut off to one piece of
equipment at a time.
• Utility service requirements are as shown on the equipment drawing.
Verify the equipment drawing or identification plate (located over
electrical box cover) for proper voltage.
2. Assure the unit is placed, as shown on equipment drawing, in correct
relation to building supply lines.
NOTE: If unit is not at the installation site, refer to Uncrating Instructions (see
Section 2.1.2) for proper moving instructions.
3. Verify that the bottle filler is level.
a. Place a spirit level end-to-end and side-to-side on top of filling area.
b. If necessary, level the unit end-to-end and side-to-side, by adjusting the
leveling legs.
4. Unit is installed on Non-slip floor surface.
These specifications are intended to describe the technical information given
on the nameplate of your unit and to state other relevant information. Verify
Equipment Drawing or nameplate, for proper voltage and amperage.
• Voltage, Amperage, and Power Consumption
The
Basil 1000 Bottle Filler
operates on:
• 120 V~, 60 Hz, 1-phase, 2-wire
• Maximum current and power consumptions are:
• 120 V~, 10A, 1.2 kW
• Permissible Environmental Conditions
This equipment is designed to give optimal results in an indoor environment
where temperature is maintained between 41-104°F (5-40°C) and where
maximum relative humidity is 80% for temperatures up to 88°F (31°C),
decreasing linearly to 50% relative humidity at 104°F (40°C).
• Seismic Anchorage System
A Seismic Tie Down System is available for high risk seismic zones.
2.2 Installation
Instructions
2.2.1 Before Installing
Equipment
2.2.2 Technical
Specifications
Figure 2-2. Disconnect
Switch
Summary of Contents for Basil 1100
Page 1: ...OPERATOR MANUAL Basil 1100 Feeder Bottle Filler 2004 03 04 P122996 004 ...
Page 2: ......
Page 4: ...ii 122996 004 Operator Manual Table of Contents THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 22: ...3 6 122996 004 Operator Manual Component Identification THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 28: ...4 6 122996 004 Operator Manual Operating Instructions THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 34: ...5 6 122996 004 Operator Manual Routine Maintenance THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 42: ...7 6 122996 004 Operator Manual Illustrated Parts Breakdown Figure 7 3 Roller Assemblies ...
Page 44: ...7 8 122996 004 Operator Manual Illustrated Parts Breakdown THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 48: ......