2.3
INSTRUMENT CARE
If no more experiments are to be performed that day, or if there was a spill on the
instrument, clean the instrument as detailed in Section 5 – Care and Maintenance.
2.3.1
Maintenance Tab
To conserve energy, turn off the instrument at the end of the work day.
x
Navigate to the “Instrument Tasks” section of the graphical user interface
(Figure 20).
x
Select “Basic Shutdown Protocol” from the “Maintenance” tab.
x
Wait until the screen becomes black, and then turn the switch on the right
side of the instrument to the off position.
Two other protocols can be run from this page: “Basic prime” and “Home all.” If an
issue arises, STEMCELL Technologies Technical Support may ask that these
protocols be run and the results observed in order to assist with troubleshooting.
Note: The Service software and Package for Diagnosis can be launched from this window, allowing for calibration adjustment and
diagnostics. This is not necessary unless requested by a STEMCELL Technologies representative.
2.3.2
Aseptic Operation of RoboSep
™
To ensure aseptic operation, RoboSep™ must be used in a biocontainment hood. RoboSep™ is designed to fit easily in most
commercially available hoods. In addition, RoboSep™ does not affect sterility within the biocontainment hood; this has been verified by
particle count as per ISO standard 14644-1.
Although the hydraulic fluid is internal to the instrument and does not come into contact with the sample, reagents, or tips during
normal operation, the use of sterilized deionized water is recommended for aseptic operation.
Standard RoboSep™ protocols dispense fluids into the sample liquid to ensure aerosol containment. Aerosol-resistant filter tips are
supplied with each RoboSep™ kit.
Figure 20 – The shutdown protocol, located under
Maintenance on the Instrument Tasks tab