Chapter 4
Programming 129
SR865A DSP Lock-in Amplifier
Example
SCNAMPATTN 1
Set the scan amplitude attenuator mode to fixed
SCNAMPATTN?
Returns the mode for amplitude output attenuators
SCNDCATTN(?) {i}
The
SCNDCATTN i
command sets the operating mode for the output attenuators when
scanning dc level to automatic (i=0) or fixed (i=1).
Normally, the sine out dc level hardware will automatically switch one or more balanced
attenuators into the dc level signal path to optimize performace for the commanded level.
During scans, however, it can be helpful to force the attenuators to remain in a fixed
configuration, inhibiting the relay switching that would otherwise introduce transients in
the sine output.
Example
SCNDCATTN 1
Set the scan dc level attenuator mode to fixed
SCNDCATTN?
Returns the mode for dc level output attenuators
SCNINRVL(?) {i}
The
SCNINRVL i
command sets the parameter update interval according to the table
below.
i Update
Interval
i Update
interval
0
8 ms
9
9.375 s
1
16 ms
10
28.12 s
2
31 ms
11
56.25 s
3
78 ms
12
112.5 s
4
155 ms
13
5 m 37 s
5
469 ms
14
11 m 15 s
6
938 ms
15
22 m 30 s
7
1.875 s
16
45 m 00 s
8 4.688
s
The parameter update interval is the approximate time spent at each scan step along the
way during a scan.
Example
SCNINRVL 8
Set parameter update interval to 4.688 s
SCNINRVL?
Returns parameter update interval index i
SCNENBL(?) { OFF | ON | i }
The
SCNENBL i
command turns scanning off (i=0) or on (i=1). This sets the scan
parameter to its begin value but does not start a scan.
Example
SCNENBL ON
Turn scanning on
Summary of Contents for SR865A
Page 5: ...Safety and Preparation For Use iii SR865A DSP Lock in Amplifier...
Page 6: ...iv Safety and Preparation For Use SR865A DSP Lock in Amplifier...
Page 54: ...36 Getting Started Chapter 1 SR865A DSP Lock in Amplifier...
Page 118: ......
Page 172: ......
Page 186: ...168 The FFT Display Appendix B SR865A DSP Lock in Amplifier...
Page 192: ......