112 Programming
Chapter
4
SR865A DSP Lock-in Amplifier
ICUR(?) { 1MEG | 100MEG | i }
The
ICUR i
command sets the current input gain to 1
MΩ
(1 μA)
(i=0) or 100
MΩ
(10 nA) (i=1).
Example
ICUR 100MEG
Set the current input gain to 100 MΩ
(10 nA)
ICUR 1
ICUR?
Returns the current input gain i
ILVL?
The
ILVL?
query returns the signal strength indicator from lowest (0) to overload (4).
Example
ILVL?
Returns the signal strength indication
SCAL(?) { i }
The
SCAL i
command sets the sensitivity according to the table below.
i sensitivity
i sensitivity
0 1
V
[
μA
] 15
10 μV
[pA]
1
500 mV [nA]
16
5
μ
V [pA]
2
200 mV [nA]
17
2
μ
V [pA]
3
100 mV [nA]
18
1
μ
V [pA]
4
50 mV [nA]
19
500 nV [fA]
5
20 mV [nA]
20
200 nV [fA]
6
10 mV [nA]
21
100 nV [fA]
7
5 mV [nA]
22
50 nV [fA]
8
2 mV [nA]
23
20 nV [fA]
9
1 mV [nA]
24
10 nV [fA]
10 500
μ
V [pA]
25
5 nV [fA]
11 200
μ
V [pA]
26
2 nV [fA]
12 100
μ
V [pA]
27
1 nV [fA]
13
50 μV
[pA]
14
20 μV
[pA]
Example
SCAL 6
Set the sensitivity to 10 mV [nA]
SCAL? Returns
the
sensitivity
i
Summary of Contents for SR865A
Page 5: ...Safety and Preparation For Use iii SR865A DSP Lock in Amplifier...
Page 6: ...iv Safety and Preparation For Use SR865A DSP Lock in Amplifier...
Page 54: ...36 Getting Started Chapter 1 SR865A DSP Lock in Amplifier...
Page 118: ......
Page 172: ......
Page 186: ...168 The FFT Display Appendix B SR865A DSP Lock in Amplifier...
Page 192: ......