118153-001 Rev. B
40
uX_Manual.doc REV -
6.12
Request
Status
Description:
The host requests that the firmware sends the power supply status.
Direction:
Host to supply
Syntax:
<STX><22><,><CSUM><ETX>
Example:
<STX>22,<CSUM><ETX>
Response:
<STX><22><,><ARG1><,><ARG2><,><ARG3><,><CSUM><ETX>
Where:
<ARG1> 1 = HvOn, 0 = HvOff
<ARG2> 1 = Interlock Open, 0 = Interlock Closed
<ARG3> 1 = Fault Condition, 0 = No Fault
Example:
<STX>22,1,1,1,<CSUM><ETX>
Notes:
The fault indicator indicates 1 of 3 possible faults.
- Interlock Fault, Interlock Opened while High Voltage was On
- Over-Voltage Fault, Output Voltage exceeds 106% of unit maximum.
- Configuration Fault, Unit Configuration stored in NV memory is invalid.
This response will also be sent in an unsolicited manner when
either the interlock is opened while High Voltage is On (interlock fault), or
an Overvoltage fault occurs. This unsolicited message is sent once when
the fault occurs, with the fault argument set to 1. Thereafter, the fault will
again read as 0. To get detailed fault status use the Request Expanded
Status Command.