118153-001 Rev. B
10
3.5 USB Cabling In High EMI Environments
If the uX USB interface is being used in a high-EMI environment, ferrites should
be added to the USB cable. The figure below illustrates the possible
combinations of ferrites that can be used to achieve acceptable operation under
these conditions.
Block Diagram if USB Cable Utilizing Ferrites
Ferrite beads should be attached to the USB cable next to the connectors – both
sides should be installed. In extreme cases ferrite cores may be added where the
cable is looped 3 or 4 times around the core as shown in figure 11. Cores of 1.5 to 2
inches should be used at both ends of the cable. In addition, a ferrite core may be
required on the 24VDC input.