118153-001 Rev. B
37
uX_Manual.doc REV -
6.9 Request Filament Current Limit
Description:
The host requests that the firmware report the filament current limit
setpoint. The A/bit scaling for the argument is model dependent. See the
tables in section 8 of this document.
Direction:
Host to supply
Syntax:
<STX><17><,><CSUM><ETX>
Response:
<STX><17><,><ARG><,><CSUM><ETX>
Where:
<ARG> = 0 - 4095 in ASCII format
Example:
<STX>17,4095,<CSUM><ETX>